1.
Int. braz. j. urol ; 31(5): 470-471, Sept.-Oct. 2005. ilus
Article in English | LILACS | ID: lil-418167

ABSTRACT

Percutaneous tube nephrostomy (PTN) placement is associated with bleeding complications in a small proportion of cases. We study a case of inadvertent renal vein catheterization during PTN tube change with catheter right atrial migration treated by fluoroscopically monitored catheter removal.


Subject(s)
Female , Humans , Middle Aged , /instrumentation , Catheters, Indwelling/adverse effects , Foreign-Body Migration , Heart Atria , Nephrostomy, Percutaneous/instrumentation , /adverse effects , Device Removal/methods , Nephrostomy, Percutaneous/adverse effects